place an order Place an Order
contact us Contact Us
request more information Request Pricing/More Info

  Products
  Training

  Tech Support
    Service Overview
    Calibration
    Warranty
    Product Registration
    Extended Service Plans
    Rentals
    Contact Tech Support

  Tech Support FAQs:
    Phoenix® 6.0 / EV30
    Phoenix®(Classic)
    FC10 / FC10Plus
    FC20
    PBA

Complete Ordering Information
Mouthpieces
Printing
Calibration Equip.
DOT Supplies
Accessories

Supplies by Tester:
Phoenix® 6.0/EV30
Phoenix® (Classic)
FC Series
Phoenix® 6.0/EV30
FC Series
Software Pkgs.
Buyer's Guide
Product FAQs
Lease-Purchase

Call today 800.722.4872
click to go to home page
Go to:  Go to Law Enforcement Go to Corrections Testing Go to Workplace Testing Go to Schools Testing Go to International Testing  
   PBA Frequently Asked Technical Questions

My PBA says “Log Full”. How do I clear this message?
What batteries should I use in my PBA?
How often should I calibrate my PBA?
How often should I send in my PBA for service?

How do I change the time on my PBA?
My PBA scrolls “Temperature out of Range” when attempting to calibrate or verify.


My PBA says “Log Full”. How do I clear this message?
Pressing buttons 3 & 4 simultaneously will clear the log and allow you to perform further testing.

<top of page>

What batteries should I use in my PBA?
The PBA uses 8 AA cell alkaline batteries. Energizer® brand are recommended.

<top of page>

How often should I calibrate my PBA?
The PBA should be calibrated every 6 months. In addition, check with your program administrator for any additional requirements or guidelines your organization may have.

<top of page>

How often should I send in my PBA for service?
The PBA should be sent to the factory for service once a year.

<top of page>

How do I change the time on my PBA?
To change the time in your PBA, press buttons 2 & 4 simultaneously, the display will flash “Ready” briefly and return the PBA 3000 screen. Press buttons 1 thru 4 sequentially, the display will scroll “Set Options Config”. Press button 3 and the time will display with the hours blinking. Press button 1 to increase, button 2 to decrease and button 3 to change fields. Once the desired time is set, press button 4 to save.

<top of page>

My PBA scrolls “Temperature out of Range” when attempting to calibrate or verify.
The PBA needs to be at an internal temperature between 68 and 86 degrees. It should give its current temperature after scrolling the above message. If it is below 68 degrees then the unit must be warmed. If it is above 86 degrees, then it must be cooled. If the message persists, contact Lifeloc Tech Support for further assistance.

<top of page>